Impeller Flow Sensors
250 and 228 Metal Tee
Installation & Operation Manual
872021-EN (August 2012)
Rev. 8
IMPORTANT:
This manual contains important information.
READ AND KEEP FOR REFERENCE.
Vue de la page 0
1 2 3 4 5 6 ... 15 16

Résumé du contenu

Page 1 - Impeller Flow Sensors

Impeller Flow Sensors250 and 228 Metal TeeInstallation & Operation Manual872021-EN (August 2012)Rev. 8IMPORTANT: This manual contains important in

Page 2 - CONTENTS

the studs should all line up perfectly when the sighting holes are parallel to pipe. If this has been accidentally loosened, please refer to the insta

Page 3 - INTRODUCTION

SPECIFICATIONSWetted Materials (except tees) • See Technical Brief for material specifics.Sensor Sleeve and Hex Adapter for 228BR, 228CB, • Sleeve:

Page 4 - MECHANICAL INSTALLATION

Repeatability• ± 0.3% of full scale over recommended design flow range for all except 250B• ± 0.7% of full scale over recommended design flow range

Page 5 - Page 5 August 2012

TROUBLESHOOTING• If the voltage at the sensor input is less than 7.0 VDC in a no flow situation, disconnect the sensor from the barrier strip and mea

Page 6 - Page 6 August 2012

Page Left Blank Intentionally.250 and 228 Metal Tee Impeller Flow SensorsPage 14 August 2012

Page 7 - Electrical Installation

Page Left Blank Intentionally.Installation & Operation ManualPage 15 August 2012

Page 8 - CALIBRATION

www.badgermeter.comData Industrial is a registered trademark of Badger Meter, Inc. Other trademarks appearing in this document are the property of th

Page 9 - Page 9 August 2012

CONTENTSIntroduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 10 - Page 10 August 2012

INTRODUCTIONUsed in conjunction with any Badger Meter impeller flow monitor or transmitter, Badger Meter non-magnetic flow sensors provide an accurate

Page 11 - SPECIFICATIONS

Model 228CBThis model consists of a Model 220BR sleeve and hex mounting adapter assembly inserted in a 2 inch or 2-1/2 inch Class 150 cast iron tee an

Page 12 - Page 12 August 2012

To install the sensor tee1. There must be free, unrestricted pipe for at least 10 diameters upstream and 5 diameters downstream of the tee. Note: If

Page 13 - TROUBLESHOOTING

ELECTRICAL INSTALLATION  STANDARD SENSORS1. The metal collar on the top of the Series 220 sensors or an optional conduit cap on the Series 250 senso

Page 14 - Page 14 August 2012

Electrical InstallationThe Series 200 Sensor is approved, as an entity, as Intrinsically Safe when installed in conformance with Badger Meter installa

Page 15 - Page 15 August 2012

CALIBRATIONBadger Meter sensors use unique K and Offset numbers for calibration. These numbers are derived from calibration runs using NIST traceable

Page 16

IMPELLER ASSEMBLY AND SHAFT REPLACEMENTIf you are replacing an existing Badger Meter impeller or shaft and have already calibrated your fl ow monitor/t